Synchronization 7 Analog Modules
System Reference, January 2001
253
For a digitizer or sampler, you have to move the trigger
signal edge backward by the trigger line length plus the
trigger-to-signal delay from the reference point, then move
it forward by the signal line length. As a result, the digi-
tizer can start the measurement at the moment when the
analog signal arrives at the analog module pin.
Figure 119 Adjusting Synchronization Timing (4) at DUT Pin for Digitizer and
Sampler m
Trigger signal
@digital pin
Measurement start
@analog module
pogo pin
Analog signal
@analog module
pogo pin
Signal line
length
Analog signal
@DUT pin
Trigger line
length
Trigger-to-signal
delay
Trigger edge setting
Move backward by
trigger line length
Move backward by
trigger-to-signal delay
Move forward by
signal line length
Reference
Point
Trigger signal
@trigger input pin
of analog module
Comentarios a estos manuales